OLYMPUS WINTER & IBE GMBH VERSAPOINT II BIPOLAR ELECTRODE RESECTOSCOPIC - 4.0 MM ANGLED LOOP, RESECTING; WA47787A Back to Search Results
Model Number WA47787A
Medical Device Problem Codes Break (1069); Material Fragmentation (1261); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The loop broke and fell inside the patient's uterine cavity.Forceps were used to retrieve the fallen device and the therapeutic hysteroscopic myomectomy was completed using a backup device.No reports of further patient harm.
 
Additional Manufacturer Narrative
This report is related to the following linked patient identifiers: (b)(6).The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the correct h6 code and the inadvertently left-out reporter's country from the previous report.Corrected field: h6 medical device problem code from material fragmentation (a040103) to break (a0401) and detachment of device or device component (a0501).Updated field: e1 - country to oman.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
No additional information was received from the customer.
 
Event or Problem Description
No additional information received from the customer.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPOINT II BIPOLAR ELECTRODE RESECTOSCOPIC - 4.0 MM ANGLED LOOP, RESECTING
Common Device Name
WA47787A
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22903500
Report Number9610773-2025-05412
Device Sequence Number15453950
Product Code HIN
UDI-Device Identifier14042761086261
UDI-Public14042761086261
Combination Product (Y/N)N
Initial Reporter CountryMU
PMA/510(K) Number
K111751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA47787A
Device Lot NumberU2402092
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/07/2025
Supplement Date Received by Manufacturer09/10/2025
10/16/2025
Initial Report FDA Received Date08/27/2025
Supplement Report FDA Received Date09/10/2025
11/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-