| Catalog Number |
68LP-350 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: improper or incorrect procedure or method.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "filled to 460cc".This record is for the left side.Device remains implanted.
|
| |
|
Search Alerts/Recalls
|