| Catalog Number |
180551 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Erosion (1750)
|
| Date of Event |
03/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: comp rvrs 25mm bsplt ha+adptr cat# 010000589 lot# 66588047.Comp lk scr 3.5hex 4.75x20 st cat# 180551 lot# 66783779.Comp lk scr 3.5hex 4.75x15 st cat# 180550 lot# 66854061.Comp rvs cntrl 6.5x25mm st/rst cat# 115395 lot# 66758637.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The reported event is unconfirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a revision procedure approximately 6 months post implantation due to implant disassociation and massive glenoid erosion.The patient was converted to a hemiarthroplasty and will be implanted with a custom baseplate on an unknown date in the future.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Search Alerts/Recalls
|