It was reported that during placement, the foley catheter was placed on (b)(6) and naturally removed on (b)(6).Leak location was unknown.Per internal bd comments received on 22aug2025, reported that18fr silicone or latex catheter.Since the balloon did not inflate even after being naturally removed, it was possible that the balloon catheter was damaged.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that during placement, the foley catheter was placed on (b)(6) and naturally removed on 17thaugust.Leak location was unknown.Per internal bd comments received on (b)(6) 2025, reported that18fr silicone or latex catheter.Since the balloon did not inflate even after being naturally removed, it was possible that the balloon catheter was damaged.
|
|
The reported event has been confirmed and is being investigated by capas 12866756 & 12474528.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley catheter with syringe.Visual inspection noted no obvious visible observations.The balloon was inflated with 10ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water), and solution immediately leaked from a 0.013" pinhole located on the catheter balloon.This is out of specification per inspection procedure ip7603444, revision 0, which states, "catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe." risk review, labeling review, and dhr review are not required as the event is being investigated per capas 12866756 & 12474528.Corrections: d, e, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|