| Model Number |
9010000849 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from user facility via a manufacturer representative regarding instruments used for minimally invasive tran foraminal lumbar interbody fusion.It was reported that driver head snapped off.There was no patient involvement reported.There were no further complications reported regarding the event.2025 aug 12, email(rep) : the rod inserted does not lock down.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis : part # 9010000849 : lot # rs14l007 visual and functional inspection revealed the inner shaft does not move when the locking lever is moved.The c clip that connects the adjusting mechanism to the inner shaft appears to have been overloaded and broken.It is possible to overload the clip when making tension adjustments while the rod is in place.It appears the c clip broke from overload medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|