| Model Number |
9-PDA-005 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2025 an 8-6mm amplatzer duct occluder was selected for implant using a 7f non-abbott delivery sheath.During implant the occluder took on a cobra shape.The occluder was not released from the delivery cable.The occluder was removed from the patient.There were no interactions with cardiac structures.There were no angulations or kinks noted on the delivery system.There were no adverse effects to the patient.The patient status was stable.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2025 an 8-6mm amplatzer duct occluder was selected for implant using a 7f non-abbott delivery sheath.During implant the occluder took on a cobra shape.The occluder was not released from the delivery cable.The occluder was removed from the patient.There were no interactions with cardiac structures.There were no angulations or kinks noted on the delivery system.There were no adverse effects to the patient.The patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.There were no complaints associated with any other devices from the lot.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|