MAUDE Adverse Event Report: DEVILBISS HEALTHCARE DEVILBISS 5L OXYGEN CONCENTRATOR; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEVILBISS HEALTHCARE DEVILBISS 5L OXYGEN CONCENTRATOR; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 525DS
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Codes Dyspnea (1816); Syncope/Fainting (4411)
Date of Event 08/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Devilbiss healthcare was notified of an incident involving an oxygen concentrator by the end user who reported the device was not putting out enough oxygen and was alarming.The end user became short of breath and administered a nebulizer treatment.The end user called ems and stated that she "passed out." the end user was admitted to the hospital for a couple days.The end user is currently on steroids.Product labeling states," if you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek medical assistance immediately to avoid harm.As a precaution, your devilbiss provider may supply you with a reserve oxygen system.If your unit loses electrical power or fails to operate correctly, the patient alert system will sound to signal you to switch to your reserve oxygen system (if provided) and contact your devilbiss provider." the device operated to design specifications by alarming to alert the user of an event.As part of its complaint review and evaluation process, devilbiss healthcare will file an update if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEVILBISS 5L OXYGEN CONCENTRATOR
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
DEVILBISS HEALTHCARE
100 devilbiss drive
somerset PA 15501
Manufacturer (Section G)
DEVILBISS HEALTHCARE
100 devilbiss drive
somerset PA 15501
Manufacturer Contact
jess bohrer
100 devilbiss drive
somerset, PA 15501
5169984600
MDR Report Key22905204
Report Number2515872-2025-00029
Device Sequence Number8001111
Product Code CAW
UDI-Device Identifier00885304000846
UDI-Public(01)00885304000846
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K071397
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number525DS
Device Catalogue Number525DS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2025
Initial Report FDA Received Date08/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age64 YR
Patient SexFemale
-
-