| Model Number |
610-2231411-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed an "ecg monitoring failure" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including bench testing and ecg stressing, without duplicating the customer's report.The customer's multifunction cable was not returned and could not be visually inspected or functionally tested.The defib receptacle was replaced as a precaution.The device was recertified and returned to the customer.The customer was advised to discard the mfc used during the event.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|