Model Number 6900PTFX
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 08/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with a 25mm 6900ptfx mitral valve underwent a valve-in-valve procedure after an implant duration of 2 years, 10 months due to calcification and severe stenosis.The patient presented with symptoms of heart failure.The tmvr was successfully performed with a 26mm 9755rsl valve.The patient was in stable condition post procedure.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: g3, g6, h2, h6: (type of investigation, investigation findings, investigation conclusions).Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.A device history record (dhr) review was performed, and no relevant non-conformances were identified.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including sever chronic renal failure on dialysis and hyperlipidemia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT PLUS MITRAL PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key22905312
Report Number2015691-2025-07032
Device Sequence Number8851672
Product Code DYE
UDI-Device Identifier00690103163582
UDI-Public(01)00690103163582(17)260810
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/10/2026
Device Model Number6900PTFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2025
Supplement Date Received by Manufacturer09/11/2025
Initial Report FDA Received Date08/27/2025
Supplement Report FDA Received Date09/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age64 YR
Patient SexFemale
-
-