| Model Number |
VTICM5_13.2 (-9.50/1.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6-clinical code.4581: significant reduction in iridocorneal angles, significant shallowing of ac.H6-investigation type 4110: lens work order search- no similar complaint event(s) within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced an excessive vault with significant reduction of iridocorneal angles, and significant shallowing of the ac.The lens remains implanted.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
H11- disregard initial mdr as is was reported in error and n/a.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|