| Catalog Number |
CAT00223 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the transport cannula broke off in the joint.The surgeon was able to retrieve the pieces that broke off.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the transport cannula broke off in the joint.The surgeon was able to retrieve the pieces that broke off.
|
| |
|
Additional Manufacturer Narrative
|
|
The sales representative confirmed that the device will not be returned.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: cannula broke off probable root cause: application - excessive force - poor visualization through arthroscope the reported failure mode will be monitored for future reoccurrence.
|
| |
|