• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION VERIFY ENHANCED; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION VERIFY ENHANCED; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON Back to Search Results
Model Number 353101
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a manufacturer representative (rep) regarding a patient who was implanted with an implantable neurostimulator (ins) for urinary/bowel disfunction.It was reported that the rep called today in a case reporting they saw high impedances on the lead with the ens on one contact, in connecting to ins they then saw all 4 contacts have high impedances.The agent asked about checking with the ens again but they were with the ins.Asked if they had set up a program to run for a few minutes to check bellows/toes.The rep reported that the patient was not seeing any motor responses due to underlying issues.The system was only letting them go up 1.2 ma and nothing further.The agent asked to check the ins set screw, make sure lead was snug after tightening which caller stated they did.The agent reviewed trying a new ins, if results do not change, then consider lead replacement.In an email response the rep indicated that the first time, impedance failed on 0 and 1.They re-ran impedance a second time and it failed on all four electrodes.After speaking with it, the rep was directed to change out the ins, which they did.Impedance still failed.They switched out the lead and the battery.This issue was not resolved.
 
Additional Manufacturer Narrative
Section d references the main component of the system.Other medical products in use during the event include: brand name lead; product id neu_unknown_lead (lot: unknown); product type: 0200-lead; implant date ; explant date.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERIFY ENHANCED
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key22909026
Report Number2182207-2025-02207
Device Sequence Number15657024
Product Code EZW
UDI-Device Identifier00763000879273
UDI-Public(01)00763000879273(11)250426(20)01
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number353101
Device Catalogue Number353101
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2025
Supplement Date Received by Manufacturer08/28/2025
Initial Report FDA Received Date08/27/2025
Supplement Report FDA Received Date09/09/2025
Date Device Manufactured04/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexUnknown
-
-