• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS ACHIEVE®; BIOPSY NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS ACHIEVE®; BIOPSY NEEDLE Back to Search Results
Catalog Number A1411/B
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A review of the device history and complaint database could not be performed since the lot number was not provided.
 
Event or Problem Description
The account alleges that when attempting to take biopsies the biopsy device failed to collect a sample.No patient injury to report.
 
Additional Manufacturer Narrative
The suspect device was returned for evaluation.The complaint was confirmed.The root cause is likely due to improper pin and spring assembly performed by the operators during the manufacturing process.Issue-(b)(4) has been initiated for further investigation.A search of the complaint database was performed and seven similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Corrective actions are taken as warranted.Trend data will be used to monitor complaints and the performance of production processes and quickly address any deviations from established standards.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE®
Common Device Name
BIOPSY NEEDLE
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS
8830 siempre viva rd #100
san diego CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS
8830 siempre viva rd #100
san diego CA 92154
Manufacturer Contact
adrian cole
1600 w merit pkwy
south jordan, UT 84095
8012084225
MDR Report Key22910000
Report Number1721504-2025-00304
Device Sequence Number8863139
Product Code KNW
UDI-Device Identifier10885403088346
UDI-Public10885403088346
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K141552
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberA1411/B
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2025
Supplement Date Received by Manufacturer09/03/2025
Initial Report FDA Received Date08/27/2025
Supplement Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-