MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37603
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: product id 3708660 serial# (b)(6) implanted: (b)(6) 2016 explanted: (b)(6) 2025 product type extension.Section d information references the main component of the system.Other relevant device(s) are: product id: 3708660, serial/lot #: (b)(6), ubd: 15-apr-2019, udi#:(b)(4).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient underwent a battery change and an extension revision on (b)(6).The manufacturer representative stated that the patient developed an infection in the left ipg pocket, which appeared red and puffy.The physician decided to remove the left chest battery and the left-sided extension due to the infection.The representative was notified of the infection on (b)(6), and the explant procedure was performed on (b)(6).
 
Additional Manufacturer Narrative
Continuation of d10: product id 37603 serial# (b)(6) implanted: (b)(6) 2022 explanted: (b)(6) 2025 product type implantable neurostimulator.Product id 3708660 serial# (b)(6) implanted: (b)(6) 2016 explanted: (b)(6) 2025 product type extension.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information received from rep indicating cause of the revision on july 24th was impedances were detected on the extension during a battery replacement procedure due to battery depletion.Impedances were resolved with extension revision.It is unknown when infection was first observed.Patient has since had their new battery and extension placed on (b)(6) 2025.The rep is not aware of any issues with the system since new battery and extension were placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key22910222
Report Number3004209178-2025-14532
Device Sequence Number9130806
Product Code MHY
UDI-Device Identifier00763000384340
UDI-Public(01)00763000384340(17)230414
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/14/2023
Device Model Number37603
Device Catalogue Number37603
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2025
Supplement Date Received by Manufacturer12/10/2025
Initial Report FDA Received Date08/28/2025
Supplement Report FDA Received Date12/12/2025
Date Device Manufactured10/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-