• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA ATLAN A350; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA ATLAN A350; ANESTHESIA UNITS Back to Search Results
Model Number 8621500
Medical Device Problem Codes Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device posted a ventilator failure alarm during use.No injury reported.
 
Additional Manufacturer Narrative
The investigation is ongoing.Results will be provided with a separate follow-up-report.
 
Additional Manufacturer Narrative
The investigation was performed based on the given information including the electronic log file.The reported issue could be confirmed and it was determined that a failure of the optical incremental encoder was the root cause.The device detected the failure and reacted as specified for this situation.In case of such failure, atlan will still provide fresh gas (including agent if a vaporizer is connected) and monitoring allowing for manual ventilation.Based on the currently available information, the risk related to the specific situation has been initially evaluated.Since similar complaints have been reported recently, a capa (corrective and preventive actions for products) was initiated.
 
Event or Problem Description
It was reported that the device posted a ventilator failure alarm during use.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAN A350
Common Device Name
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23558
GM   23558
Manufacturer Contact
moislinger allee 53-55
lübeck 23558
4518822868
MDR Report Key22910734
Report Number9611500-2025-00481
Device Sequence Number18358610
Product Code BSZ
UDI-Device Identifier04048675556176
UDI-Public(01)04048675556176(11)230929(93)8621500-15
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K230931
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8621500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/15/2025
Supplement Date Received by Manufacturer10/10/2025
Initial Report FDA Received Date08/28/2025
Supplement Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-