| Model Number |
8621500 |
| Medical Device Problem Codes |
Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.Results will be provided with a separate follow-up-report.
|
| |
|
Event or Problem Description
|
|
It was reported that the device posted a ventilator failure alarm during use.No injury reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the device posted a ventilator failure alarm during use.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was performed based on the given information including the electronic log file.The reported issue could be confirmed and it was determined that a failure of the optical incremental encoder was the root cause.The device detected the failure and reacted as specified for this situation.In case of such failure, atlan will still provide fresh gas (including agent if a vaporizer is connected) and monitoring allowing for manual ventilation.Based on the currently available information, the risk related to the specific situation has been initially evaluated.Since similar complaints have been reported recently, a capa (corrective and preventive actions for products) was initiated.
|
| |
|
Search Alerts/Recalls
|