MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE E; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE E; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42532007102
Medical Device Problem Codes Loss of or Failure to Bond (1068); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Joint Impingement (4834); Aseptic loosening (4883)
Date of Event 06/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2- australia.H3- customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient was revised approximately two years eight months post implantation due to tibial loosening leading to anterior-medial tibial wall impingement of keel.Attempts to obtain additional information have been made; however, no more information is available at this time.
 
Additional Manufacturer Narrative
This report is being submitted to provide additional information.Updated: b4, b5, d2, g1, g3, g6, h1, h2, h3, h6, and h11 corrected: d4 (udi); h4 no product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: thin radiolucent lines around the tibial stem that can be seen with mechanical loosening or infection.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.This complaint is not confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE E
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22911399
Report Number3007963827-2025-00338
Device Sequence Number9130695
Product Code MBH
UDI-Device Identifier00889024471061
UDI-Public(01)00889024471061(17)320802(10)65433815
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number42532007102
Device Lot Number65433815
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2025
Supplement Date Received by Manufacturer10/22/2025
Initial Report FDA Received Date08/28/2025
Supplement Report FDA Received Date11/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-