| Model Number |
CCA0T0 |
| Medical Device Problem Code |
Material Discolored (1170)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The product was not returned for analysis.Only photo was returned.The reported complaint cannot be confirmed from the returned photo.Received photo/video shows what appears to be an implanted iol (intraocular lens), the reported complaint cannot be observed in the returned photo.The root cause is deemed to be manufacturing related.Complaints have been received describing that lenses were reported by doctors for the presence of a ¿polychromatic sheen¿ visible.The cause of ¿sheen¿ or ¿film-like¿ appearance is a result of a change in the positioning of the iol during the manufacturing process and is not associated with any foreign material or contamination.The observation has been reported clinically and confirmed in laboratory settings to disappear over time following implantation.The appearance is a result of a temporary change in the surface of the iol that occurs during the delivery process, which is only visible under certain orientations and illumination settings.This resolves once the iol is delivered and exposed to body temperature over time.There are no adverse impacts or lasting effects on the iol or the patient.Based on the results of the product investigation and testing, it was confirmed there are no adverse events or clinical impact on vision associated with this finding.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is:(b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported following an intraocular lens (iol) implant procedure, there was a fine film on the anterior surface of the iol.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|