• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CCA0T0
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.11.The product was not returned for analysis.Only photo was returned.The reported complaint cannot be confirmed from the returned photo.Received photo/video shows what appears to be an implanted iol (intraocular lens), the reported complaint cannot be observed in the returned photo.The root cause is deemed to be manufacturing related.Complaints have been received describing that lenses were reported by doctors for the presence of a ¿polychromatic sheen¿ visible.The cause of ¿sheen¿ or ¿film-like¿ appearance is a result of a change in the positioning of the iol during the manufacturing process and is not associated with any foreign material or contamination.The observation has been reported clinically and confirmed in laboratory settings to disappear over time following implantation.The appearance is a result of a temporary change in the surface of the iol that occurs during the delivery process, which is only visible under certain orientations and illumination settings.This resolves once the iol is delivered and exposed to body temperature over time.There are no adverse impacts or lasting effects on the iol or the patient.Based on the results of the product investigation and testing, it was confirmed there are no adverse events or clinical impact on vision associated with this finding.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is:(b)(4).
 
Event or Problem Description
A physician reported following an intraocular lens (iol) implant procedure, there was a fine film on the anterior surface of the iol.Additional information was requested.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22911479
Report Number9612169-2025-01717
Device Sequence Number21094366
Product Code HQL
UDI-Device Identifier00380652397377
UDI-Public(01)00380652397377(17)270927(11)230929(10)25726286
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCA0T0
Device Lot Number25726286
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/31/2025
Supplement Date Received by Manufacturer10/17/2025
Initial Report FDA Received Date08/28/2025
Supplement Report FDA Received Date11/11/2025
Date Device Manufactured09/29/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISCOVISC
Patient Age77 YR
Patient SexMale
-
-