• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NELLCOR PURITAN BENNETT MEXICO SA NELLCOR OXYSOFT; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NELLCOR PURITAN BENNETT MEXICO SA NELLCOR OXYSOFT; OXIMETER Back to Search Results
Model Number OXYSOFTN
Medical Device Problem Codes Incorrect Measurement (1383); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the sensor had inconsistent readings with motion and a cable breach was identified.The sensor was exposed to natural lighting, applied to the finger, and held in place without additional securing methods.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted a wrap had been pushed back and the black and red wires were visible.A comprehensive examination could not be performed, because the returned sample was not received in a state that allowed full functional or visual assessment.It was reported that the sensor had inconsistent readings with motion.The reported issue could not be confirmed.The most likely cause could not be established from the information available.It was also reported that the sensor had a cable breach.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Correction: h3.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted that the wrap had been pushed back and the black and red wires were visible.A comprehensive examination could not be performed, because the returned sample was not received in a state that allowed full functional or visual assessment.It was reported that the sensor had inconsistent readings with motion.The reported issue could not be confirmed.The most likely cause could not be established from the information available.It was also reported that a cable breach was identified.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR OXYSOFT
Common Device Name
OXIMETER
Manufacturer (Section D)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX  22225
Manufacturer (Section G)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX   22225
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22912226
Report Number1282497-2025-00565
Device Sequence Number9110805
Product Code DQA
UDI-Device Identifier10884521818972
UDI-Public(01)10884521818972(17)260630(10)231840221H
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K212555
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberOXYSOFTN
Device Catalogue NumberOXYSOFTN
Device Lot Number231840221H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2025
Supplement Date Received by Manufacturer09/01/2025
11/12/2025
12/15/2025
06/05/2026
Initial Report FDA Received Date08/28/2025
Supplement Report FDA Received Date09/05/2025
12/09/2025
12/19/2025
06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-