| Model Number |
8CN85H |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Hypoventilation (1916); Unintended Extubation (4564)
|
| Date of Event |
07/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced low exhaled tidal volume (vte) on the ventilator due to an air leak around the stoma, causing inadequate ventilation.The pilot balloon was mostly deflated, with size fluctuations corresponding to patient respirations.Minimal leak technique (mlt) was performed, but the cuff failed to hold pressure.An emergency tracheostomy tube change was performed.Post-procedure testing of the tube via a water air leak test confirmed a blown cuff.
|
| |
|