AMO MANUFACTURING NETHERLANDS TECNIS SIMPLICITY; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DRN00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 07/02/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient implanted with tecnis odyssey simplicity 16.5 d intraocular lens (iol) on (b)(6) 2025 in the right eye (od) initially reported symptoms of blurred and hazy vision, followed by difficulty reading and mild blurred vision at a distance.On (b)(6) 2025, the lens was explanted during a secondary surgery and replaced with the same model intraocular lens of 17 d.The procedure did not require unplanned incision enlargement, sutures, or vitrectomy, and there was no procedural delay.The patient¿s pre-operative best corrected visual acuity was 20/25-2, and post-operative visual acuity improved to 20/20.The patient was fully recovered.No further information is available.
 
Additional Manufacturer Narrative
Section a3b: unknown; information not provided.Section h3: other 81: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
The product was returned to the manufacturing site for evaluation.Additional information: section d9: device available for evaluation? yes section d9: date returned to manufacturer: september 10, 2025 section h3: evaluated by manufacturer: yes device evaluation: visual inspection under magnification was performed on the lens.No issues were found that could cause or contribute to the complaint issue reported were observed.No further evaluation was performed.The complaint issues reported could not be confirmed during product evaluation, and no issues were identified.Based on the complaint investigation results, the product was released within specifications.No product deficiency or product malfunction could be identified.No nonconformity report, documentation or labeling changes, and further escalations are required.Johnson & johnson surgical vision will continue to monitor these types of complaints.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key22913459
Report Number3012236936-2025-000226
Device Sequence Number13023631
Product Code MFK
UDI-Device Identifier05050474810693
UDI-Public(01)05050474810693(17)270409
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDRN00V
Device Catalogue NumberDRN00V0165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2025
Supplement Date Received by Manufacturer09/12/2025
Initial Report FDA Received Date08/28/2025
Supplement Report FDA Received Date10/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
Patient Weight58 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-