• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SILK ROAD MEDICAL ENROUTE TRANSCAROTID STENT; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SILK ROAD MEDICAL ENROUTE TRANSCAROTID STENT; STENT, CAROTID Back to Search Results
Model Number SR-1040-CS
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Ischemia Stroke (4418); Thrombosis/Thrombus (4440)
Date of Event 08/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced a thrombotic stroke two hours post procedure requiring additional intervention.An enroute transcarotid stent was selected for use in a right sided transcarotid artery revascularization (tcar) procedure.The procedure was completed with no notable issues.Two hours post procedure the patient experience stroke like symptoms.The physician explanted the stent, patched the right carotid artery and noted very small platelet aggregates at the distal end of the stent.Imaging was performed and there were no abnormalities.The patient woke up neurologically intact after the procedure with no symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID STENT
Common Device Name
STENT, CAROTID
Manufacturer (Section D)
SILK ROAD MEDICAL
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL INC.
27th ave north plymouth
minneapolis MN 55447
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key22913494
Report Number2124215-2025-59425
Device Sequence Number15749167
Product Code NIM
UDI-Device Identifier00811311020539
UDI-Public(01)00811311020539(17)270430(10)18322741
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSR-1040-CS
Device Catalogue NumberSR-1040-CS
Device Lot Number0018322741
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2025
Initial Report FDA Received Date08/28/2025
Date Device Manufactured05/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient RaceBlack Or African American
-
-