DAVIS & GECK CARIBE LTD MAXON; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number GMM-344L
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: gmm-344l, gmm-344l maxon* 1 grn 240 cm lp gs26x12 (lot# d4l2766y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during an emergency veterinary surgery, when closing the linea alba of a horse at the completion of a colic procedure.The suture broke before the needle was passed into the horse.The suture was discarded, and another packet was opened.The second packet also broke, but at the end of the suture line, when the knot was being tied.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d4 (udi), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.One opened sample with the appropriate packaging, and two sealed, representative samples were returned for evaluation.Visual inspection of the opened sample noted that the needle was disengaged from the suture.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.For the sealed samples, visual inspection could not find any abnormalities.It was reported that the suture broke before the needle was passed into the horse.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXON
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22916118
Report Number9612501-2025-02293
Device Sequence Number9865285
Product Code GAM
UDI-Device Identifier10884523004786
UDI-Public(01)10884523004786(17)291031(10)D4L2766Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K990951
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGMM-344L
Device Catalogue NumberGMM-344L
Device Lot NumberD4L2766Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2025
Supplement Date Received by Manufacturer01/16/2026
Initial Report FDA Received Date08/28/2025
Supplement Report FDA Received Date01/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-