| Catalog Number |
367336 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® push button blood collection set customer noticed leakage.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D.2b medical device type: one additional code applies; jka.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® push button blood collection set customer noticed leakage.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary : bd received one photo for investigation.The evaluation of the photo showed no evidence of leakage.The photo depicted an unused blister pack.Additionally, ten retained samples were used for functional tests, and no leakage occurred.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: leakage.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|