COVIDIEN LLC BIS; INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR Back to Search Results
Model Number 186-1046
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to use, the unit¿s screen went black and then showed incorrect readings.There was no patient involvement.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, g6, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the unit¿s screen went black and then showed incorrect readings.The reported issue was not confirmed the most likely cause could not be identified because no related problem was detected with the device.The evaluation detected an unreported condition: the front cover and back cover were both cracked/damaged.The most likely cause for the additional condition was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that prior to use, the unit¿s screen went black and then showed incorrect readings.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Common Device Name
INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR
Manufacturer (Section D)
COVIDIEN LLC
6135 gunbarrel ave
boulder,co
Manufacturer (Section G)
COVIDIEN LLC
6135 gunbarrel ave
boulder,co
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22921417
Report Number2936999-2025-00432
Device Sequence Number16040677
Product Code OLW
UDI-Device Identifier10884521129511
UDI-Public(01)10884521129511(11)220721
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K072286
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186-1046
Device Catalogue Number186-1046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2025
Supplement Date Received by Manufacturer09/09/2025
Initial Report FDA Received Date08/29/2025
Supplement Report FDA Received Date10/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-