| Model Number |
A20710A |
| Medical Device Problem Codes |
Break (1069); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the forceps jaws failure to grasp malfunction: broken jaws wire, a component failure that is expected or random without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the subject forceps exhibited a broken jaw wire.There was no patient involvement.
|
| |
|