RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number BLX3100S19
Medical Device Problem Codes Circuit Failure (1089); Electrical /Electronic Property Problem (1198); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to a ventilator inoperative condition occurring in certain bipap, and mechanical ventilator devices.The manufacturer received information alleging a ventilator inoperative condition on the device.The device was returned to a third-party service center for evaluation and the customer's complaint was confirmed.The device¿s airpath kit which contains the main pca and blower was replaced to address the issue.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.This issue has been identified under the fsn 2023-cc-src-039.Additionally, an alarm indicating a coin cell voltage is out of range was observed in the device's downloaded event log.The device's main pca replacement addressed this issue.This would not affect the device's ability to deliver therapy as designed.The device was tested and passed all final testing.The bipap a40 pro (blx3100s19) is substantially similar to the bipap a40 (1111169) and will be reported in the united states under bipap a40, 501k number: k121623.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22922285
Report Number2518422-2025-109504
Device Sequence Number9079419
Product Code MNT
UDI-Device Identifier00606959059122
UDI-Public(01)00606959059122
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 08/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberBLX3100S19
Device Catalogue NumberBLX3100S19
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Initial Report FDA Received Date08/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/28/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-