| Catalog Number |
07.02030.003 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that he central screw section of the adj crosslink was damaged.During the final tightening process, the threads were stripped, preventing proper fastening, and metal fragments were also observed.There was no patient harm or procedural impact.
|
| |
|
Event or Problem Description
|
|
It was reported that he central screw section of the adj crosslink was damaged.During the final tightening process, the threads were stripped, preventing proper fastening, and metal fragments were also observed.There was no patient harm or procedural impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d9 and h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the center screw has deformation damage.Root cause: this event could possibly be attributed to an improperly sized driver being used to tighten the cross connector screw, resulting in stripping of the screw head and shearing of material.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|