|
(b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on (b)(6) 2025.An investigation was conducted on (b)(6) 2025.The loading device was returned taken apart with the cover of the loading device off the base of the device as well as the blue wings not inside the loading device.Signs of clinical use and evidence of blood was observed on the loading device.The delivery device was returned outside the loading device with the white plunger not depressed and the blue safety lock on, which prevents the white plunger from being depressed.The seal and tension spring assembly was observed outside the loading device.The seal was detached from the tether with the blue string intact.No other visual defects were observed.Measurements of the delivery device were taken; the inner diameter was measured at 0.195 inches; the outer diameter was measured at 0.219 inches.The length of the delivery tube was measured at 2.50 inches.The measurement values recorded for the delivery tube were within the tolerance specifications.Based on the returned condition of the device as well as the evaluation results, the reported failure "fitting problem" was not confirmed, however, the analyzed failure "detachment of device or device component" was observed.The lot # 3000474241 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|