| Catalog Number |
0275628000 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the small joint shaver released two small shards of metal into the wrist joint.Doctor was able to remove both pieces prior to closing.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the device was not saved.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: device breaking.Probable root cause: inadequate window profile, inadequate welding process, improper material strength, inadequate size selected for the application, material brittleness, excessive force applied by the user, incorrect rfid tag.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the small joint shaver released two small shards of metal into the wrist joint.Doctor was able to remove both pieces prior to closing.
|
| |
|
Search Alerts/Recalls
|