Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735821r, serial/lot #: (b)(6), ubd: udi#: (b)(4), h3: a manufacturer representative went to the site to test the equipment.In summary, the camera was replaced.Codes fdm b01, fdr c08, fdc d02 are applicable to this analysis.D9, h3: the hardware was returned and analysis was performed.The returned positioning sensor unit (psu) had scratches on the housing and lenses.Event logs reported intermittent firmware incompatibility dating back to sep 2020.There was a battery voltage low message along with bump detected and storage temperature exceeded.The psu failed an accuracy test (aak) at.641mm with a passing threshold of.250mm.Codes fdm b01, fdr c02, fdc d02 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|