| Model Number |
ABSTACK30X |
| Medical Device Problem Codes |
Positioning Failure (1158); Separation Problem (4043)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the laparoscopic hernia repair, after mesh placement, the tack was not fired properly and would not hold the mesh well.The surgeon used a suture to fix the mesh to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: e1 (first name, last name), g3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d4, d9, g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection and functional testing found no notable condition.It was reported that after mesh placement, the tack was not fired properly and would not hold the mesh well.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition of bent shaft.The product analysis noted evidence that the device was not used as intended.This issue may occur if the instrument is applied with excessive force or side load, causing excess torque on the rotating helices and tube shaft which can cause poor tack penetration.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: appropriate force should be applied to the handle of the device when placing tacks.Excessive force may result in damage to the tissue, device and/or the material being fixated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|