• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB BASE UNIT SERVO-N; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB BASE UNIT SERVO-N; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 6688600
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while the ventilator was ventilating a patient in the niv nava (non-invasive ventilation neurally adjusted ventilatory assist) mode of ventilation, a leakage was detected.There was no patient harm.Manufacturer's ref.#: (b)(4).
 
Event or Problem Description
Manufacturer's ref.#: (b)(4).
 
Additional Manufacturer Narrative
The investigation is based on the received logs and analysis performed by our subject matter expert and tests were performed in the laboratory.Review of the received logs revealed a significant number of check tubing alarm activations can be found on (b)(6) 2025.Check tubing alarm may be triggered in very high leakage situations with at least some patient interfaces with high expiratory resistance due to significant pressure difference between the inspiratory and expiratory pressure transducer.This in turn causes the safety valve and expiratory valve to open for 5 s, resulting in loss of peep (positive end expiratory pressure).The technical log did not contain any technical error codes that could indicate a malfunction of the ventilator.The most common triggering criteria for the check tubing alarm is the exp pressure variation.Breaking up the circuit at the y-piece will with almost 100 % certainty result in check tubing triggered by the exp pressure variation during insp being too small.Our conclusion is that there is no indication of a ventilator malfunction at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE UNIT SERVO-N
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key22932250
Report Number8010042-2025-0001345
Device Sequence Number10231413
Product Code CBK
UDI-Device Identifier07325710001042
UDI-Public(01)07325710001042(11)191203
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K201874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6688600
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2025
Supplement Date Received by Manufacturer10/28/2025
Initial Report FDA Received Date09/02/2025
Supplement Report FDA Received Date10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/22/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-