| Model Number |
E161 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.To date, this icd remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.To date, this icd remains in service.No adverse patient effects were reported.According to the additional information, this icd was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|