DEPUY SPINE INC IS ROD SIDE CUTTER; INSTRUMENT, CUTTING, ORTHOPEDIC Back to Search Results
Catalog Number 205027
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that two fine tipped berbecker rod cutters had been chipped due to use in the or over the past several weeks.There were no patient consequences.Procedure was completed successfully.It is unknown on what date the damage occurred.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b3.Corrected (g3) alert date.The correct alert date for this complaint is (b)(6) 2025 based on aei note.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history review (dhr): a review of the receiving inspection (ri) for is rod side cutter, was conducted identifying that lot number hm402223, was released in one batch.¿ batch1: lot qty of (b)(4) units were released on dec 21, 2022, with no discrepancies.Supplier: (b)(4).The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IS ROD SIDE CUTTER
Common Device Name
INSTRUMENT, CUTTING, ORTHOPEDIC
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key22936541
Report Number1526439-2025-01110
Device Sequence Number18592057
Product Code HTZ
UDI-Device Identifier10705034001258
UDI-Public(01)10705034001258(10)HM402223
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number205027
Device Lot NumberHM402223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/12/2025
Supplement Date Received by Manufacturer11/04/2025
Initial Report FDA Received Date09/02/2025
Supplement Report FDA Received Date11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-