| Model Number |
G5S-80C |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device displayed a red "x" indicator.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the device activity logs makes it clear the device presented an error in (b)(6) 2023 where the end result was that the device code indicator was red and was beeping.The device will communicate to the user that it is not rescue ready by providing an auditory beep every 30 seconds and displaying a red rescue ready (nvi) indicator.We can identify from the log that this device appears to have been left idle in a non-usable state for at least 2 years prior to reporting it to zoll.Zoll medical corporation evaluated the device.The device was put through extensive testing including daily/weekly/monthly testing, on/off current testing, on/off current testing, and defibrillation cycling without duplicating the report.The batteries were not returned for evaluation.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|