• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 15 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 15 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 15-265
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported left side rupture.The device has been explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: d9, h3, h6 device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed an opening assessed as fold flaw opening.As per the investigation procedure, creases and stress marks were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
 
Event or Problem Description
Healthcare professional reported left side rupture.The device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key22937515
Report Number9617229-2025-14820
Device Sequence Number9819747
Product Code FTR
UDI-Device Identifier10888628001435
UDI-Public(01)10888628001435(10)2661595(11)141010(17)190910
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/10/2019
Device Catalogue Number15-265
Device Lot Number2661595
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/06/2025
Supplement Date Received by Manufacturer10/09/2025
Initial Report FDA Received Date09/02/2025
Supplement Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-