| Catalog Number |
SRLP-220 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported ¿rupture¿.This record is for the right side.Device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a4.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿rupture.¿ this record is for the right side.Device was explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿rupture¿.This record is for the right side.Device was explanted and replaced.Device was returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed multiple broken devices through microscopic inspection 1 broken device assessed as fold flaw opening and inconclusive and 1 opening assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (stress marks) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|