| Catalog Number |
133S-MX-12-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "ruptured tissue expander".This record is for right side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: deflation: observed an opening assessed as surgical damage.As per the investigation procedure, opening is at the suture tab were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "ruptured tissue expander".This record is for right side.The device was explanted.
|
| |
|