• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A30; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A30; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number 1116155
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
A bipap a30, silver series device was returned to a third-party service center for remediation.There was no injury or harm reported.The device was not reported to be in patient use.During the evaluation of the device, at a third-party service center the device was visually inspected and found no evidence of foam degradation.The customer complaint was not confirmed.Additionally, a critical error code e-47 and e-96 was noted to have occurred indicating the central processing unit cannot communicate with the pressure sensor using spi bus.Replacement of the device's system board was required to address the issue.The device passed all final testing.This device has been serviced, inspected, tested, met acceptance criteria in accordance with all of the applicable customer requirements as defined in the contractual agreement with plexus.The device history record has been reviewed and approved by plexus quality assurance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A30
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22939576
Report Number2518422-2025-109580
Device Sequence Number21298331
Product Code MNS
UDI-Device Identifier00606959039384
UDI-Public(01)00606959039384
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K113053
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1116155
Device Catalogue Number1116155
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2025
Initial Report FDA Received Date09/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/18/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-