PARKER MEDICAL PARKER FLEX-TIP; ENDOTRACHEAL TUBE, HIGH VOLUME LOW PRESSURE (PFHV), CUFFED, PARKER, 7.0MM
|
Back to Search Results |
|
| Model Number |
H-PFHV-70 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is not possible as no lot number was provided.The documented udi is based on the stock/product code provided by the reporter; the udi-pi is not available as no lot number was provided.All information reasonably known as of 02 sep 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported: after the operation was completed, the medical staff noticed that air leak had occurred, the patient was not reintubated.There was no harm or injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information-investigation completion: h2; h6.Functional testing was performed on the returned product, by the distributor; the test results confirmed there was an air leak in the cuff; the reported issue, of air leak was confirmed; however, the root cause was not determined.A risk assessment was performed, and the ultimate risk was determined to be low.A review of the device history record is not possible as no lot number was provided.The documented udi is based on the stock/product code provided by the reporter; the udi-pi is not available as no lot number was provided all information reasonably known as of 04 mar 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported: after the operation was completed, the medical staff noticed that air leak had occurred, the patient was not reintubated.There was no harm or injury reported.
|
| |
|
Search Alerts/Recalls
|
|
|