| Model Number |
8884710859E |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported they opened up two different tubes and the stylet was stuck on both.They placed the ng tube, flushed the tube using a syringe, and tried removing the stylet.They could not get any of the stylets removed and had to send the patient to radiology to get the tube placed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Prior to a lot¿s release, the lot must be deemed acceptable by passing inspections that are based on a valid sampling plan.During production, inspectors routinely examine a statistical sample both physically and visually.A device was not returned for evaluation; therefore, the affected product could not be evaluated to confirm the reported failure mode.As such, a root cause could not be determined.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|