| Catalog Number |
21-7092-24 |
| Medical Device Problem Codes |
Disconnection (1171); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had a faulty patient-controlled analgesia (pca) connection resulting in leakage and faulty patient-controlled analgesia (pca) connection broke apart.There was no patient injury.The event occurred at the hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
D9 - date returned to mfg: 10/7/2025.H3 and h6 codes - updated.Two used devices were returned for investigation.The devices were visually inspected.One of the samples had the asv luer taper broken off and stuck inside of the y-site.Both samples had dried, sticky residue at the connection between the asv valve and the y-site.During leak test, the customer reported issue was confirmed as a leak was observed between the asv valve and the y-site.For the broken sample, the luer stuck in the y-site was plugged with a pin gage and red dye was inserted through the tubing.No leaks were observed from the bond site of the asv luer and the y-site.The probable cause is due to excessive twisting/bending force on the asv luer.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|
| |
|