| Model Number |
470007-05 |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the customer experienced an issue with an intuitive product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The round tip scissors instrument was analyzed and found to have a frayed pitch cable at the distal end.Some filaments of the cable were frayed, but the cable is not fully broken.The frayed cable strands stuck out at the wrist.There were no evidence of discoloration or corrosion/contamination on the cables to indicate a reprocessing induced issue.Further inspection found no abnormally sharp or damaged components that could have contributed to the cable breaking.Field h10 is blank as there are no related report numbers.
|
| |
|