It was reported that the patient was experiencing pain in his leg and felt that something had changed since his surgery.X-rays revealed that the distal tip of the stem had fracture off and was lose inside the canal.Surgeon removed ceramic head, removed the old stem which came out easy and the trephinex down to remove the small piece that was lodged distal.In addition, the surgeon swapped out the poly liner for a new one, cemented in a new stem and head before reducing.No further issues were noted.Doi: (b)(6) 2024, dor: (b)(6) 2025, affected side: left hip.
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: h3, h6: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|