H11.Additional narrative/data the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
It was reported that a patient with a 25mm 7300tfx mitral valve is under evaluation for a valve-in-vale procedure after an implant duration of three (3) years and six (6) months due to unknown reason.The patient presented heart failure, cough, shortness of breath, fatigue, and leg swelling.
|
|
H11: additional manufacturer narrative: updated sections g3, g6, h2, h6 (investigation findings, investigation conclusions).Bioprosthetic tissue valves can deteriorate with time and eventually fail contributing to regurgitation and/or stenosis.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular regurgitation and/or stenosis.Due diligence attempts were made to gather additional information regarding a device failure mode.However, additional information is not available at this time.In this case, there is no evidence of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, a definitive root cause cannot be conclusively determined.Per the instructions for use (ifu), reoperation/explant are known potential adverse events associated with bioprosthetic cardiac valves and annuloplasty rings.The device history record (dhr) was not reviewed as no information regarding a device failure mode was provided.Therefore, a review of manufacturing and/or device component issues that may have contributed to the reported event is unable to be performed.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.If new information becomes available, a supplemental report will be submitted.
|