MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) AVEIR¿ RETRIEVAL SYSTEM CATHETER; DEVICE, PERCUTANEOUS RETRIEVAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) AVEIR¿ RETRIEVAL SYSTEM CATHETER; DEVICE, PERCUTANEOUS RETRIEVAL Back to Search Results
Model Number LSCR111
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented in clinic on (b)(6) 2025 for initial implant procedure.During procedure, it was found that the right ventricular (rv) leadless pacemaker (lp) was dislodged during tether mode.When the rvlp was being recovered while it was still connected to the delivery catheter, the rvlp separated prematurely from the catheter, causing embolism, and the rvlp was sitting freely in the right ventricle.When the rvlp was being retrieved by the retrieval catheter, the catheter was loosed up to re-adjust the device for retrieval, but the rvlp slipped and migrated into the right atrium.Eventually, the rvlp was retrieved, not implanted, and an alternate near at hand was implanted successfully instead.The first delivery catheter used was then examined, where it was found that activation failure was exhibited.Patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEIR¿ RETRIEVAL SYSTEM CATHETER
Common Device Name
DEVICE, PERCUTANEOUS RETRIEVAL
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key22948708
Report Number2017865-2025-99162
Device Sequence Number8259907
Product Code MMX
UDI-Device Identifier05415067034496
UDI-Public(01)05415067034496(10)10400078(17)250831
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K213494
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model NumberLSCR111
Device Lot Number10400078
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2025
Supplement Date Received by Manufacturer09/03/2025
Initial Report FDA Received Date09/02/2025
Supplement Report FDA Received Date09/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight89 KG
-
-