MAUDE Adverse Event Report: ARTHREX, INC. BTB TIGHTROPE; SMOOTH FIXATION PIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BTB TIGHTROPE; SMOOTH FIXATION PIN Back to Search Results
Model Number BTB TIGHTROPE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2025, it was reported by an arthrex employee via email that for an ar-1588btb, btb tightrope®, after ar-1588rt-ib broke (case-(b)(4)), ar-1588btb was opened to be used as a new femoral tightrope.When cinching the graft into the femoral socket, the btb tightrope broke.It was also reported that for an ar-2324pslc, suture anchor, peek-swivelock® c, 4.75 x 19.1 mm, vented, when attempting to secure the internal brace to the shin, the anchor threads broke/bent upon insertion.The doctor drilled 4.5mm into the medial tibia and used 4.75 swivelock tap.This was detected during use in a acl graftlink procedure on (b)(6) 2025.The procedure was completed successfully as a 2nd ar-1588rt-ib tightrope was opened and used successfully.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Complaint allegation is confirmed by the attached picture, which shows that the white suture was cut.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude the most likely cause.The most likely cause of the reported failure can be attributed to user error, due to improper surgical technique, and/or excessive force used during suture passing or the device coming into contact with another device during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BTB TIGHTROPE
Common Device Name
SMOOTH FIXATION PIN
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22950923
Report Number1220246-2025-03701
Device Sequence Number9636595
Product Code MBI
UDI-Device Identifier00888867019515
UDI-Public(01)00888867019515(11)210216(17)260131(10)12739588
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K241235
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberBTB TIGHTROPE
Device Catalogue NumberAR-1588BTB
Device Lot Number12739588
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2025
Supplement Date Received by Manufacturer11/13/2025
Initial Report FDA Received Date09/03/2025
Supplement Report FDA Received Date11/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-