|
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.During the procedure, when suturing blood vessels, one side of needle was separated from thread.(during using on the patient) all impact products were discarded.No complaint sample can be returned.No adverse patient consequences were reported.Additional information was requested.
|
|
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: ¿ did this issue contribute to any patient adverse event? >> no, hcp changed a new suture to complete the wound closure.Additional h11: was surgery delayed due to the reported event? -- unknown, was procedure successfully completed? -- unknown, were fragments generated? -- unknown, if yes, were they removed easily without additional intervention? -- unknown, patient status/ outcome / consequences -- , was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: -- unknown, is the patient part of a clinical study -- unknown, (b)(4).Device property of -- none, device in possession of -- none.
|