| Model Number |
GF-UCT180 |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the evis exera ultrasound gastrovideoscope tested positive for an unexpected contamination.The issue was found during a routine culture of the scope.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide corrections and additional information based on the completed investigation's review of the customer cleaning disinfection and sterilization (cds) processes, the device evaluation and the final investigation.Updated fields: d8; d9; e1; g3; h2; h3; h6 and h11.Corrected field: e1.The device was culture tested positive by the customer.In addition, the device evaluation found the adhesive on a (angulation)-rubber was detached.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The user request was not confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause was not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|