MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that the evis exera ultrasound gastrovideoscope tested positive for an unexpected contamination.The issue was found during a routine culture of the scope.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Event or Problem Description
No additional information received from the customer.
 
Additional Manufacturer Narrative
This report is being supplemented to provide corrections and additional information based on the completed investigation's review of the customer cleaning disinfection and sterilization (cds) processes, the device evaluation and the final investigation.Updated fields: d8; d9; e1; g3; h2; h3; h6 and h11.Corrected field: e1.The device was culture tested positive by the customer.In addition, the device evaluation found the adhesive on a (angulation)-rubber was detached.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The user request was not confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause was not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Common Device Name
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22952353
Report Number3002808148-2025-14527
Device Sequence Number18550897
Product Code ODG
UDI-Device Identifier04953170341793
UDI-Public04953170341793
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K093395
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/06/2025
Supplement Date Received by Manufacturer10/02/2025
Initial Report FDA Received Date09/03/2025
Supplement Report FDA Received Date10/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1